MSCGUIDE (Sep98) mscred MSCGUIDE (Sep98)
GUIDE TO THE NOAO MOSAIC DATA HANDLING SOFTWARE
Francisco Valdes
September 1998
MSCRED Version 2.0
SECTIONS
1. Introduction
2. Multiextension FITS Files
3. Examining Mosaic Data
3.1 Displaying the Data
2.1.1 On-the-Fly (OTF) Calibration
2.1.2 Real-Time Display with the DCA
3.2 Examining the Data
3.3 Examining the Headers
3.4 Determining Best Focus
4. Data Reductions
4.1 Some Preliminaries
4.2 Basic CCD Calibration
4.2.1 Calibration Data to Obtain At the Telescope
4.2.2 Preparing Calibration Data
4.2.3 Pupil Image Removal from Flat Fields
4.2.3.1 Broadband Data
4.2.3.2 Narrowband Data
4.2.4 Object Exposure Reductions
4.2.5 Pupil Image Removal from Object Data
4.2.5.1 Broadband Data
4.2.5.2 Narrowband Data
4.2.6 Dark Sky or Twilight Sky Flat Fields
4.2.7 The Variable Pixel Scale and Zero Point Uniformity
4.3 Coordinate Calibration
4.3.1 Setting Coordinate Zero Points and Measuring Coordinates
4.3.2 Matching Coordinate Systems
4.4 Putting the Pieces Together
4.4.1 Removing Sky Gradients
4.4.2 Constructing Single Images
4.4.3 Matching Intensity Scales
4.4.4 Making the Final Stack Image
1. INTRODUCTION
This document discusses handling and reducing CCD mosaic data,
particularly data from the NOAO CCD Mosaic Imager (referred to here as
the NOAO Mosaic), using IRAF and the MSCRED package. It is not a
beginner's guide and assumes some previous experience with IRAF and CCD
reductions.
The first section discusses the mosaic data format and how to use it
with IRAF. This format is more complex than single CCD images because
of the multiple CCDs and possibly multiple amplifiers per CCD. To keep
the data from each exposure self-contained the multiple CCD images are
stored in a single file. This multiple image per file has many
advantages but it does mean that some commands for dealing with images
behave differently.
The second section describes the tools used to examine the mosaic data.
These tools are used during observing as well as during data reductions.
The last section describes the reduction of mosaic data. This includes
basic CCD instrumental calibration and combining mosaic exposures into
single images.
2. MULTIEXTENSION FITS FILES
The data format used by the NOAO Mosaic Data Handling Software (MDHS)
is a multiextension FITS (MEF) file. This format is produced by the
the Data Capture Agent (DCA) when observing with the NOAO Mosaic. The
MEF file for the NOAO Mosaic currently consists of nine FITS header and
data units (HDU). The first HDU, called the primary or global header
unit, contains only header information which is common to all the CCD
images. The remaining eight HDUs, called extensions, contain the
images from the eight CCDs.
The fact that the image data is stored as a FITS file is not
significant. Starting with IRAF V2.11, FITS files consisting of just a
single primary image may be used in the same way as any other IRAF
image format. The significant feature of the mosaic format is its
multi-image structure.
With multiextension FITS files you must either use tasks which are
specifically designed to operate on these files as a unit or explicitly
specify the image within the file that is to be operated upon by general
IRAF image processing tasks. The tasks in the MSCRED package are
designed to operate on the mosaic MEF files and so you only need to
specify the filename. For image tasks outside the MSCRED package you
must specify the image in the MEF file using the syntax
filename[extension]
where "filename" is the name of the MEF file. The ".fits" filename
extension is optional provided there is no confusion with other files
with the same basename. The image "extension" is specified either
using an extension name or the position of the extension in the file
(where the first extension is 1). The extension names in the NOAO
Mosaic data are "im1" through "im8" for the eight CCDs. For a detail
discussion of the IRAF FITS Image Kernel and the syntax it supports for
multiextension FITS files see
ftp://iraf.noao.edu/iraf/docs/fits_userguide.ps.Z.
If you forget to specify an extension to a task that expects only
single images you will get the following error which is your reminder
to include an extension.
ms> imhead obj012 1
ERROR: FXF: must specify which FITS extension (obj012)
Two of the most common tasks that require specifying an image extension
are DISPLAY to display a single CCD image (the task MSCDISPLAY is used
to display all the images at once) and IMHEADER to list the header of a
particular CCD. So, for example, the following commands might be used.
ms> display obj012[im2] 1
ms> imhead obj012[3] l+
Other tasks you may use this way are IMEXAM and IMPLOT.
A common question is how to specify a list of extensions. Modification
of the syntax to allow wildcard templates in the extension
specification is under study. Currently you must specify each
extension explicitly, though the filename itself may be a wildcard; for
example the first image in a set of files can be collectively specified
with
obj*[im1]
There are two methods for specifying some or all extensions in tasks
that operate upon lists of images. One is to make @files. This can be
done explicitly with an editor. However the PROTO task IMEXTENSIONS
can expand MEF files into an @file as in the following example.
ms> imexten obj012,obj13 > list
ms> imhead @list
Read the help page for further information, additional parameters, and
examples.
Another method is to use the special MSCRED task MSCCMD. This task can
be used on the command line or as a simple interactive command
interpreter. The idea is that you use the special designations "$input"
and "$output" for task parameters which allow lists of images. Then
lists of MEF filenames are specified for the input and output which are
expanded and substituted into the task parameters when it is executed.
For example,
ms> msccmd "imhead $input l+" input=obj012,obj013
For additional information and examples consult the help page for that
task.
Note that the tasks IMSTAT and IMARITH are so useful and common that
there are specific MSCRED tasks MSCSTAT and MSCARITH that operate on
all or a subset of image extensions. So these tasks need not be used
with MSCCMD or with @files.
We conclude with a discussion of the special operations of copying,
renaming, deleting, and reading and writing FITS tapes as they apply to
the mosaic MEF files. To copy a mosaic file as a unit use COPY, making
sure to explicitly specify the "fits" extension. If you use IMCOPY it
will expect you to specify a particular extension and will copy only
that extension. While IMCOPY is not the way to copy an complete MEF
file the tasks IMRENAME and IMDELETE are the commands for renaming and
deleting these files; though RENAME and DELETE will also work provided
you are explicit with the extension. Finally the mosaic data should be
kept as a MEF file and so the special mosaic tasks MSCWFITS and
MSCRFITS should be used. The current WFITS and RFITS are not intended
for this type of data.
3. EXAMINING MOSAIC DATA
During observing a small set of IRAF commands are commonly used to
examine the data. This section describes these commands. While the
discussion is oriented towards examining the data at the telescope
during the course of observing, the tools described here are also used
when reducing data at a later time.
3.1 DISPLAYING THE DATA
The two commands DISPLAY and MSCDISPLAY are used to display the data in
a display server window. The display server is a separate process
which must be running before displaying the images. The observing
environment at the telescope will generally have the XIMTOOL display
server already running with a window on a separate monitor. If it is
not running for some reason it can be started with a menu selection.
Away from the telescope you would start XIMTOOL or SAOIMAGE as you do
normally.
The display server must be told what size "frame buffer" to allocate for
holding the display pixels. This determines how many pixels may be
loaded at one time. Note that the display window may be smaller than
this size and the display server allows you to move the portion viewed
and zoom/unzoom any region. If the image size is larger than the frame
buffer you can display a portion of the image at full resolution or the
full image at a lower resolution. The frame buffer size is queried and
set with the commands:
ms> show stdimage
imt4096
ms> set stdimage=imt2048
There are trade-offs in the frame buffer selection. A large frame
buffer allows you to have higher resolution for the large mosaic images
but it uses more memory and takes longer to load.
The DISPLAY task is used to display individual images in the display
server. This task is a standard IRAF task about which you are assumed
to have some basic knowledge. There are many display options which are
discussed in the help page. The only special factor in using this task
with mosaic data is that you must specify which CCD image to display
using the image extension syntax discussed previously. As an example,
to display the central portion of extension im3 in the first frame and
the whole image in the second frame:
ms> display obj123[im3] 1 fill-
ms> display obj123[im3] 2 fill+
The MSCDISPLAY task is based on DISPLAY with a number of specialized
enhancements for displaying mosaic data. It displays the entire mosaic
observation in a single frame by "filling" each image in a tiled region
of the frame buffer. The default filling (defined by the order
parameter) subsamples the image by uniform integer steps to fit the
tile and then replicates pixels to scale to the full tile size. The
resolution is set by the frame buffer size. As mentioned before,
trying to increase the resolution with a larger buffer size has the
penalty of longer display times. An example display command is:
ms> mscdisplay obj123 1
The default parameters for MSCDISPLAY are shown below. Many of the
parameters are the same as DISPLAY but there are also a few that are
specific to the task of displaying a mosaic of CCD images as indicated
with an asterisk.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mscdisplay
image = root name for image to be displayed
frame = 1 frame to be written into
* (mimpars= ) mosaic image parameters
* (check = no) check if image is loaded
* (onepass= no) load all extensions in one pass?
(bpmask = BPM) bad pixel mask
(bpdispl= none) bad pixel display (none|overlay|interpolate)
(bpcolor= red) bad pixel colors
(overlay= ) overlay mask
(ocolors= green) overlay colors
(erase = yes) erase frame
(border_= no) erase unfilled area of window
(select_= yes) display frame being loaded
(repeat = no) repeat previous display parameters
(fill = no) scale image to fit display window
(zscale = yes) display range of greylevels near median
(contras= 0.25) contrast adjustment for zscale algorithm
(zrange = yes) display full image intensity range
(zmask = ) sample mask
* (zcombin= auto) Algorithm for combining z1 and z2 values...
(nsample= 1000) maximum number of sample pixels to use
(order = 0) spatial interpolator order (0=replicate,...
(z1 = 0.) minimum greylevel to be displayed
(z2 = 1000.) maximum greylevel to be displayed
(ztrans = linear) greylevel transformation (linear|log|none|user)
(lutfile= ) file containing user defined look up table
The mapping of the pixel values to grey levels includes the same
automatic or range scaling algorithms as in DISPLAY. This is done for
each image in the mosaic separately. The new parameter "zcombine" then
selects whether to display each image with it's own independent display
range ("none") or to combine the display ranges into a single display
range based on the minimum and maximum values ("minmax"), the average
of the minimum and maximum values ("average"), or the median ("median")
of the minimum and maximum values. The independent scaling is most
appropriate for raw data while the "minmax" scaling is recommend for
processed data which are gain calibrated. The special value "auto"
(the default) checks if the display data has been flat fielded, either
by separate processing or with on-the-fly calibration, and if so it
uses "minmax" scaling and if not it used independent scaling.
The "mimpars" (mosaic image parameters) parameter is actually a
reference to another set of parameters. The default with no value is
to use the parameters from the parameter task MIMPARS. These
parameters can be examined and set with EPAR either by typing ":e"
when over this parameter in MSCDISPLAY or by running EPAR directly on
this task; i.e. epar mimpars. The parameters for NOAO Mosaic data are
shown below.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mimpars
(extname= ) extension name pattern
(exttmpl= _![1-9]![1-9]![1-9].*) extension template
(xgap = 72) minimum X gap between images
(ygap = 36) minimum Y gap between images
(process= no) do calibration processing?
(oversca= yes) do line-by-line overscan subtraction?
(flatfie= yes) do flat field correction?
(caldir = mscdb$noao/kpno/4meter/caldir/) calibration directory
(filter = !filter) filter
The "extname" parameter is used to select as subset of the image
extensions to display. It is a pattern to match extension image
names. For extensions such as im1, im2, etc. the pattern typically
uses the character selection template such as "[1256]" to select
anything with a 1, 2, 5, or 6 in the name. The pattern matching syntax
can be found in the help for the task MATCH.
The "exttmplt" parameter is for use with non-MEF data. The gap
parameters define the gap size in the display. The remaining
parameters are for the on-the-fly calibration discussed below.
3.1.1 ON-THE-FLY (OTF) CALIBRATION
Raw mosaic data can exhibit significant instrumental artifacts which may
interfere with inspecting the data prior to reductions. The most
significant artifact is gain variations both within each CCD image and
between the CCDs. In the simplest case of constant gain variations
between the CCDs the independent display scaling, "zcombine" of none or
auto, may be sufficient. But when there are significant flat field
patterns it may be desirable to apply a quick, approximate flat field
calibration as the data are being displayed.
MSCDISPLAY can apply an on-the-fly (OTF) calibration to raw mosaic
exposures. This does not change the actual data files and the
calibration is intended to be quick and approximate. The calibration
steps performed are a line-by-line bias subtraction using the overscan
region of the data and a division by a flat field. If the data have
been overscan corrected or flat field corrected by CCDPROC then the
task will automatically skip those steps. The title of the display
will indicate if the data have been calibrated by adding "[bias]" for
bias subtraction and "[bias,flat=XXX]" for bias subtraction and flat
fielding using an OTF flat field called XXX.
The bias subtraction is performed by averaging the overscan pixels in a
line and subtracting this average from all the pixels in the line. This
removes the amplifier bias and line-by-line patterns.
The flat field or response calibration is performed by reading special
compact flat field calibration data which provides an approximate
relative response for each pixel in each amplifier readout. Depending
on how the calibration file is derived this will approximately correct
for pixel sensitivity variations, gain variations between the
amplifiers, sky illumination variations, and any pupil ghost pattern
(as occurs with NOAO Mosaic data from the Mayall (KPNO 4meter)
telescope).
The "process" parameter in the MIMPARS parameter set shown earlier
selects whether to turn on or off the OTF processing. If it is no then
regardless of the "overscan" or "flatfield" parameter settings no
calibration is applied. If it is yes then one or both calibration
operations can be selected. Because the MIMPARS parameters can be set
on the command line, it is common to leave the "process" parameter set
one way, say to "no", and then override the value when displaying. For
example,
ms> mscdisplay obj023 1 proc+
ms> mscdisplay flat022 2 proc+ flatfield-
The flat field calibration requires special calibration files. The
"caldir" parameter defines a directory containing the calibration
files. This can be a standard directory or a user directory. Note
that if a directory is specified it must end with $ or /.
Within the calibration directory the calibration file to apply is
selected by the "filter" parameter. For automatic selection of
calibrations, the calibrations can be selected by the filter string in
the header (or by giving the same filter string in the "filter"
parameter). To use the filter string in the header the value of the
filter parameter is set to "!<keyword>" where <keyword> is the keyword
for the filter string.
Creating the a calibration directory and calibration files is done with
the task MSCOTFFLAT. For the NOAO Mosaic a calibration directory is
provided. However you can create your own as described in the help for
MSCOTFFLAT. The "filter" parameter can be set to one of these names.
3.1.2 REAL-TIME DISPLAY WITH THE DCA
During data acquisition the MSCDISPLAY task can be used to display
mosaic data as it is being written to disk by the DCA. It begins
execution shortly after the readout begins and displays the portion of
the recorded image which has been written to disk. It then continually
displays new data which has been written by the DCA until the exposure
is completely written to the display.
The DCA control panel allows you to select whether to display the data
during readout and how it is to be displayed. This includes selecting
the OTF calibration. One toggle is equivalent to the "process"
parameter. If the processing is turned on the DCA automatically
selects only overscan bias subtraction for non-object exposures and
selects both bias subtraction and flat field division for object
exposures. The "filter" parameter is set by passing through the filter
string from the data acquisition system or by overriding this and using
the filter menu to select one of the available calibrations.
3.2 EXAMINING THE DATA
The task MSCEXAMINE allows interactive examination of mosaic images.
It is essentially the same as the standard IMEXAMINE task except that
it translates the cursor position in a tiled mosaic display into the
image coordinates of the appropriate extension image. Line and column
plots also piece together the extensions at the particular line or
column of the mosaic display. To enter the task after displaying an
image the command is:
ms> mscexam
As with IMEXAMINE, one may specify the mosaic MEF filename to be
examined and if it is not currently displayed it will be displayed
using the current parameters of MSCDISPLAY.
It is important to realize that this task shares the MIMPARS parameters
with MSCDISPLAY. To get data values back that match what is displayed
the parameters must agree with those used to display the data. In
particular, if the data are display with OTF processing then MSCEXAM
must be told this either by explicitly setting the process flat in
MIMPARS or setting it on the command line,
ms> mscexam proc+
3.3 EXAMINING THE HEADERS
There was discussion earlier concerning the use of generic image tasks
with the NOAO Mosaic data. The tasks IMHEADER and HSELECT fall into
this category. The two important points to keep in mind are that you
must specify either an extension name or the extension position and
that the headers of an extension are the combination of the global
header and the extension headers.
Often one does not need to list all the headers for all the extensions.
The image title and many keywords of interest are common to all the
extensions. Thus one of the following commands will be sufficient to
get header information about an exposure or set of exposures:
ms> imhead obj*[1] l- # Title listing
ms> imhead obj123[1] l+ | page # Paged long listing
ms> hselect obj*[1] $I,filter,exptime,obstime yes
If you need to list header information from all the extensions then you
need to take the additional step of creating an @file or using MSCCMD.
For example to get the default read noise and gain values for each CCD:
ms> imextensions obj123 > list123
ms> hselect @list123 $I,rdnoise,gain yes
or
ms> msccmd "hselect $input $I,rdnoise,gain yes" input=obj123
The CCDLIST task in the MSCRED package is specialized for the mosaic
data. It provides a compact description of the name, title, pixel
type, filter, amplifier, and processing flags. The "extname" parameter
may be used to select a particular extension, a set of extensions, or
all extensions. Because all extensions should generally be at the same
state of reduction it may be desirable to list only the first
extension. Like most of the CCD reduction tasks you can also select
only a certain type of exposure for listing. Examples of the two modes
are:
# Summary for all exposures
ms> ccdlist *.fits extname=im1
# Summary for all object exposures
ms> ccdlist *.fits extname=im1 ccdtype=object
# List of all extensions.
ms> ccdlist obj123 extname=""
3.4 DETERMINING BEST FOCUS
Focus sequence frames can be evaluated for the best focus using MSCEXAM
and the 'r' or 'a' keys. However, there is a special task for measuring
the sequence of focus images called MSCFOCUS. This displays a focus
exposure with MSCDISPLAY (if needed) and then lets you select one or
more bright stars to measure. This task is customized so that all you
need do is mark the top image in any CCD. For NOAO Mosaic data, header
information tells the task how many exposures, the spacings between the
exposures, and the focus values. After the measurements are made they
are displayed and analyzed graphically and written to the terminal and
logfile. This task is the mosaic analog of the KPNOFOCUS and STARFOCUS
tasks for single CCD data.
4. DATA REDUCTIONS
The reduction of CCD mosaic data can be divided into two stages. The
first is the basic calibration of the individual CCDs. This stage is
similar to reducing data from single CCD exposures except that the
calibration operations are repeated for all the CCDs in the mosaic. The
only significant difference is that any scaling of an exposure, such as
in normalizing the flat field calibration, must be done uniformly over
all the CCDs. The details of repeating the calibrations for all CCDs
and the scaling of the calibration data are taken care of by the
software and the data format so that these operations appear the same
as with single CCD data.
There are some steps which are not typical for CCD data with smaller
fields of view or specific to the NOAO Mosaic at the Mayall telescope.
At the Mayall telescope there are reflections off the corrector that
produce a visible image of the pupil. Coating of the corrector
minimizes this image but it may be desirable to remove this
instrumental signature which would otherwise cause a small variation of
the photometric zero point as well as an unwanted visible feature.
There are two sections discussing removal of this feature from the flat
field data and from the object exposures. If your data is from the
KPNO 0.9 meter telescope or the image is faint enough that it is not of
concern then you can skip the extended discussion.
A caveat about the pupil removal steps described here is that this
document was written prior to the latest removal of the corrector for
better anti-reflection coating. So the NOAO staff have little
experience with these corrections though earlier work has shown that
these steps will do a good job.
Another step of the basic CCD calibration stage which has generally been
ignored or forgotten with smaller single CCD formats is the variable
pixel scale. The large field of view provided by a mosaic and the
optics required to provide it can lead to a significant variation in
the pixel scale. This effect is important with the Mayall telescope
and is also present in the NOAO 0.9 meter data to a smaller degree. It
is likely to be present in other telescopes as well.
When the pixel scale varies significantly the standard flat field
calibration operation will cause the photometric zero point to vary. A
simple calibration step can be performed to remove this effect.
However, if you intend to produce single images from the mosaic of CCDs
this step is not necessary since the resampling operation naturally
accounts for this effect.
The second stage of data reductions is unique to mosaic data. This
stage is the combining of the multiple CCD images and multiple
exposures into a single image. Since creating a single image from a
single mosaic exposure is of marginal value, the thrust of this stage
of the reductions is the combining of multiple exposures which have
been spatially offset or "dithered" to cover both the gaps between the
individual CCDs and any defects.
The steps required to produce a single deep integration from dithered
exposures consist of accurately registering the images, mosaicing the
exposures into single images with the same spatial sampling, measuring
changes in the intensity scale due to variations in transparency and sky
brightness, and combining the individual images into a single deep image
with the gaps and bad pixels removed.
4.1 SOME PRELIMINARIES
The command SETINSTRUMENT is used to set default parameters for the
tasks in the MSCRED package appropriate to a particular instrument.
For users of the NOAO Mosaic it is recommended you run this command the
first time you reduce data. Subsequently you should not do this since
it will reset parameters you later changed.
To set the parameters for reducing the NOAO Mosaic data type the command
ms> setinstrument kpno 4meter CCDMosaThin1 review-
Substitute "36inch" for "4meter" if the data is from the Kitt Peak 0.9
meter telescope.
For some of the operations it is useful to specify lists of exposures
corresponding to a dither set. The examples in this guide show using
@files for dither sets. An @file is simply a list of filenames. These
can be created in several ways including using a text editor. One way
is with the FILES command to expand a file template. For example,
ms> files obj021,obj022,obj023,obj024,obj025 > field1
ms> dir @field1
obj021 obj002 obj003 obj004 obj005
4.2 BASIC CCD CALIBRATION
Basic CCD instrumental calibrations consist of correcting each CCD for
electronic bias levels, zero exposure patterns, dark counts, and pixel
sensitivities. A cosmetic replacement of bad pixels may also be
included. For the Mayall telescope the pupil image due to reflections
off the corrector must be removed from the flat field and object
exposures. An additional calibration is required to correct for the
variable pixel scale across the field of view if you intend to do
photometry on the individual CCD images.
4.2.1 CALIBRATION DATA TO OBTAIN AT THE TELESCOPE
Good data reductions begin with obtaining good calibration data at the
telescope. This section discusses the NOAO Mosaic but the general
principles will apply to other detectors, though the relative
importance of different calibrations will depend on the quality of the
CCDs and the stability of the camera.
The standard calibration data are sequences of zero exposures and
sequences of dome flat field exposures. While dark count exposures,
matched to the typical object exposure times, were important for the
first generation (engineering grade) NOAO Mosaic, dark counts are
expected to be low in the science grade detectors. Thus dark count
exposures are probably not necessary.
Dome flat fields (dome flats) provide a fair basic flattening of the
data to 2% or so, but sky flat fields (sky flats) are required to
produce dithered data that can be combined without introducing obvious
artifacts. Good sky flats can flatten the data to 0.1%. In our
experience twilight exposures do not work well. Instead dark sky flat
fields are derived from unregisted object exposures taken during the
night or run. If your observing program consists of only large
extended objects or single pointings then you should also take some
dithered exposures of "blank" sky.
At the Mayall telescope there is a pupil image caused by reflections
off the corrector. For broadband photometry the effects of the pupil
image are small but they can be reduced even further by reduction steps
to remove the image. One useful calibration for this removal is a
narrowband dome flat field. The idea is that the narrowband flat field
has a more prominent pupil image that can be used as a template for the
much fainter broadband pupil image.
Lastly, good astrometry is required to register and stack the Mosaic
images. The NOAO Mosaic data contains previously determined astrometry
recorded in the headers of the raw exposures. This is sufficient for
most purposes. However, for cameras without astrometry or to generate
your own astrometry solutions, fields with a reasonable density of
stars with cataloged accurate coordinates must be taken. Note that
with the new generation of large astrometric catalogs and the large
field of view of a mosaic, it may be that the object exposures already
contain sufficient information for deriving new astrometric
calibrations or corrections. Note that this guide does not yet discuss
how to create the astrometric coordinate system solutions.
4.2.2 PREPARING CALIBRATION DATA
This section describes how to prepare the basic calibration data. The
steps are virtually the same as with the CCDRED package and, in fact,
the command names and parameters are the same. The basic calibration
data of zero level, dark count, and dome flat fields are generally
taken as a sequence of identical exposures which are combined to
minimize the noise. A later section discusses preparing a sky flat
field calibration using the object exposures.
The calibration exposures are individually reduced by CCDPROC and then
combined. Thus, it is necessary to first set the CCDPROC parameters.
Because this task knows which operations are appropriate for particular
types of calibration exposures you can set all the parameters for
object exposures. Below is a typical set of parameters. The main
optional setting is whether or not to replace bad pixels by
interpolation, which is purely a cosmetic correction. However, it is
recommended that this be done to avoid possible arithmetic problems in
the processing.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = ccdproc
images = List of Mosaic CCD images to process
(output = ) List of output processed images
(ccdtype= object) CCD image type to process
(noproc = no) List processing steps only?
(oversca= yes) Apply overscan strip correction?
(trim = yes) Trim the image?
(fixpix = yes) Apply bad pixel mask correction?
(zerocor= yes) Apply zero level correction?
(darkcor= no) Apply dark count correction?
(flatcor= yes) Apply flat field correction?
(sflatco= no) Apply sky flat field correction?
(biassec= !biassec) Overscan strip image section
(trimsec= !trimsec) Trim data section
(fixfile= BPM) List of bad pixel masks
(zero = Zero) List of zero level calibration images
(dark = Dark) List of dark count calibration images
(flat = Flat*) List of flat field images
(sflat = Sflat*) List of secondary flat field images
(minrepl= 1.) Minimum flat field value
(interac= no) Fit overscan interactively?
(functio= legendre) Fitting function
(order = 1) Number of polynomial terms or spline pieces
(sample = *) Sample points to fit
(naverag= 1) Number of sample points to combine
(niterat= 1) Number of rejection iterations
(low_rej= 3.) Low sigma rejection factor
(high_re= 3.) High sigma rejection factor
(grow = 0.) Rejection growing radius
The overscan correction has two methods as selected by the fitting
function. A value of "legendre" (or "chebyshev" or "spline3") take all
the overscan data and fit a smooth function along the column
direction. The "order" value of 1 shown above fits a single constant
value. This leaves to the zero level calibration to subtract any
details of line-by-line structure. A value of "mean", "median", or
"minmax" take the mean, median, or mean excluding the minimum and
maximum values, of the overscan at each line and subtract that value
from that line. The other fitting parameters are ignored. The
advantage of this is that systematic line-by-line patterns are
subtracted. The disadvantage is, since the sample of overscan pixels
is small at each line, that this can also introduce a statistical
line-by-line pattern. There is currently no recommendation for the
NOAO Mosaic.
The first step is generally to process and combine sequences of zero,
dark, and dome flat exposures. This is done using the tasks
ZEROCOMBINE, DARKCOMBINE, and FLATCOMBINE. The combining must be done
in the following order since the processing of later calibration data
requires the preceding calibration data.
ms> zerocombine *.fits
ms> darkcombine *.fits
ms> flatcombine *.fits
Each of these tasks search all the exposures for a particular type so
it is fine to specify all files, though if the file names code the
type, such as "dflatNNN", then one can use that as the wildcard to
shorten the search of all the data. Also FLATCOMBINE has the feature
that it will combine the data separately for each filter. However, you
can use explicit file lists, templates, or @files to limit the input
files. The output combined names have standard default values which
the above settings for CCDPROC use.
It is a good idea to first check that the different calibration types
and filters are correctly identified by the software. This is done
using the CCDLIST command
ms> ccdlist *.fits
Unless you change the parameters "mscred.backup" and "mscred.bkuproot"
the original raw files will be saved in the subdirectory "Raw/". If
you want to start over, delete the processed files and copy the raw
files back to the working directory. If disk space is a concern and
you are satisfied with the combined calibration files you can delete
the individual processed calibration files. There is a parameter in
the combining tasks that will delete the individual files automatically
after processing and combining.
4.2.3 PUPIL IMAGE REMOVAL FROM FLAT FIELDS
NOAO Mosaic data taken at the Mayall (4meter) telescope include a pupil
image caused by reflections off the corrector. The magnitude of this
image is a function of the filter and the state of the anti-reflection
coatings on the corrector. It is also a function of the total light,
including from outside the field of view, and somewhat on the location
of bright stars.
It might appear at first that one simply divides the object exposures by
the flat field as is done for the OTF display calibration. However
this is not photometrically correct because the pupil image is an
additive light effect and not a detector response. Instead the pupil
image must first be removed from the flat field before applying it to
the object data. The object data is then corrected after flat fielding
by subtracting the extra light from the pupil image.
The pupil image is removed from the flat field by dividing by an
estimate of the pupil image pattern. The challenge is to determine the
pupil image contribution in the presence of other flat field structure.
There are two current approaches to obtaining the pupil image pattern
for removal from the data. One is to use data from another source
where the pupil pattern is more easily separated from the flat field
pattern. The second is to derive the pattern from the data assuming
something about the form of the pattern. In particular, to use the
difference in scales between the larger pupil pattern and the smaller
flat field pattern. The first approach is preferable since it better
preserves fine structure in the pupil image but the second is needed
when no other data is available.
4.2.3.1 BROADBAND DATA
For broadband data the recommended procedure is to obtain a narrowband
flat field exposure. This narrowband exposure will have a stronger
pupil image relative to the flat field pattern and, when the pupil image
is scaled down to match the broadband image flat field, the errors from
the flat field response will be diminished.
The pupil image is extracted from the narrowband flat field using the
task MSCPUPIL. This task determines the background levels in a ring
inside and outside the main pupil image and subtracts this background to
produced the pupil image template. Outside the outer background ring
the template is set to zero. In effect this is like "scrapping off"
the pupil image from the exposure.
The relevant parameters are
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mscpupil
input = List of input images
output = List of output images
(masks = BPM) List of masks
(type = data) Output type
(xc = 27.) Pattern center offset (pixels)
(yc = 9.) Pattern center offset (pixels)
(rin = 300.) Radius of inner background ring (pixels)
(drin = 20.) Width of inner background ring (pixels)
(rout = 1500.) Radius of outer background ring (pixels)
(drout = 20.) Width of outer background ring (pixels)
(funcin =chebyshev) Inner azimuthal background fitting function
(orderin= 2) Inner azimuthal background fitting order
(funcout= spline3) Outer azimuthal background fitting function
(orderou= 2) Outer azimuthal background fitting order
* (rfuncti= spline3) Radial profile fitting function
* (rorder = 40) Radial profile fitting order
* (abin = 0.) Azimuthal bin (deg)
* (astep = 0.) Azimuthal step (deg)
(niterat= 3) Number of rejection iterations
(lreject= 3.) Low rejection rms factor
(hreject= 3.) High rejection rms factor
(datamin= INDEF) Minimum good data value
(datamax= INDEF) Maximum good data value
(verbose= yes) Print information?
The output type is set to "data" to extract the pupil image after
background subtraction. The pattern center parameters are offsets from
the astrometric center and the inner and outer radii are measured from
the pattern center. The default values are for the last measured Mayall
pupil image. The fitting parameters marked with an asterisk are not
used when extracting the pupil image directly.
The pupil image template is scaled and removed from the flat field
using the task RMPUPIL. The removal is done with the arithmetic
operation
I(out) = I(in) / (scale * I(template) + 1)
where I(out) are the output corrected pixel values, I(in) are the input
pixel values, I(template) are the pupil image template pixel values, and
scale is the relative scale factor to be applied. The parameters for
the pupil image removal task are
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = rmpupil
input = Input mosaic exposure
output = Output mosaic exposure
template= Template mosaic exposure
(type = ratio) Type of removal
(extname= [2367]) Extensions for fit
(blkavg = 8) Block average factor
(fudge = 1.6) Fudge factor
(interac= yes) Interactive?
(mscexam= no) Examine corrections with MSCEXAM?
The "input" is the broadband flat field, the "output" is the corrected
flat field, and the "template" is the narrowband pupil image produced by
MSCPUPIL. The type of removal for a flat field is "ratio" as given by
the equation above.
Determining the optimal scaling of the template pupil image to the input
pupil image is normally done interactively. The task makes a guess at
scaling. If this task is used non-interactively this will be the scale
used. When the task is used interactively the input and corrected
mosaic exposures are displayed and then a query for a new scale is
given. By repeatedly adjusting the scale factor the best visual
removal can be obtained. When done the output corrected flat field is
created using the last specified scale factor. Note that to quit
requires entering dummy special values for the scale factor. A value
of zero means to create the final output exposure with the last scale
factor and a value of -1 means to quit without producing any output.
Because this operation is fairly slow and iterative there are some steps
that can be taken to it speed up. The "extname" parameter selects just
those extensions to look at. For NOAO Mosaic data the default selects
the central four extensions covered by the pupil image. The "blkavg"
parameter applies a block average to the input exposure and template.
This makes the display and iterative corrections faster. When the best
scale factor has been determined the entire input image at full
resolution is corrected by the full resolution template to create the
output flat field. If one wants to use the facilities of MSCEXAM to
evaluate each iterative correction then the "mscexam" parameter can be
set. However, the most powerful estimate for the optimal scale factor
is viewing the display and possibly blinking between the uncorrected
and corrected frames.
4.2.3.2 NARROWBAND DATA
For narrowband data the pupil image template must be derived from the
data itself. This is done by fitting the data with an axially symmetric
pattern. The fitting is performed by MSCPUPIL with the parameters
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mscpupil
input = List of input images
output = List of output images
(masks = BPM) List of masks
(type = ratio) Output type
(xc = 27.) Pattern center offset (pixels)
(yc = 9.) Pattern center offset (pixels)
(rin = 300.) Radius of inner background ring (pixels)
(drin = 20.) Width of inner background ring (pixels)
(rout = 1500.) Radius of outer background ring (pixels)
(drout = 20.) Width of outer background ring (pixels)
(funcin =chebyshev) Inner azimuthal background fitting function
(orderin= 2) Inner azimuthal background fitting order
(funcout= spline3) Outer azimuthal background fitting function
(orderou= 2) Outer azimuthal background fitting order
(rfuncti= spline3) Radial profile fitting function
(rorder = 40) Radial profile fitting order
(abin = 0.) Azimuthal bin (deg)
(astep = 0.) Azimuthal step (deg)
(niterat= 3) Number of rejection iterations
(lreject= 3.) Low rejection rms factor
(hreject= 3.) High rejection rms factor
(datamin= INDEF) Minimum good data value
(datamax= INDEF) Maximum good data value
(verbose= yes) Print information?
Note that this only differs from the previously shown parameters by
setting the "type" parameter to ratio. Because the template is derived
from the data itself there is no need to use RMPUPIL to iteratively
determine a scale factor. The "output" parameter is the corrected flat
field.
The corrected narrowband flat field will show some artifacts from fine
structure in the pupil image. However, a large fraction of the pupil
image will be removed. Later reduction steps of applying a sky flat
field and combining with dithering further eliminate effects of this
approximate solution to the pupil image.
4.2.4 OBJECT EXPOSURE REDUCTIONS
At this point you will have some subset of combined zero level, dark
count, and flat field calibration data. The calibration data is
applied to the object exposures, either in bulk or as observations are
completed, using the task CCDPROC. The command is simply
ms> ccdproc <files>
4.2.5 PUPIL IMAGE REMOVAL FROM OBJECT DATA
The pupil ring image in the object exposures is removed by subtraction
since this is excess light. Again this is only required for data where
the pupil image occurs, such as from the Mayall telescope. The tasks
for modeling and removing the image are the same as for removal from
the flat field except that the "type" parameter is set to "difference".
4.2.5.1 BROADBAND DATA
Probably the best subtraction will be obtained by using the pupil image
template from a narrowband flat field. This would be the same as used
for the flat field and extracted from the narrowband flat field using
MSCPUPIL with "type = data". The subtraction is carried out using
RMPUPIL with "type = difference".
An alternative, since the pupil image is weak and the fine structure is
unimportant, is to use MSCPUPIL with "type = difference" to determine a
smooth large scale ring pattern and subtract it from the data. The
iterative sigma rejection and the "datamin" and "datamax" parameters are
used to eliminate smaller scale astronomical objects in the field from
affecting the background fits and the ring profile fits. For this
application the "abin" parameter should be set to a value such as 30
degrees and the "astep" parameter to a smaller value such as 5 degrees.
The main advantage of this method is that no iterative scaling is
required since the fit is done directly to the data. The difficulty,
though, is if there is a bright star or fairly extended object,
particularly in the inner background ring, then the fit will be poor
and the subtraction will show gross artifacts.
The last alternative, and the one to use if there is no narrowband flat
field for the template and the field has bright stars which affect
fitting directly to the data, is to make a "sky flat" to generate the
pupil image template. This is done as described in the section for
creating a sky flat. Once the sky flat is created with the pupil image
then MSCPUPIL is used to separate the pupil image from the background
and RMPUPIL is used to scale and subtract the image from the object
exposures. Note that after the pupil image is subtracted then a new
sky flat should be created.
4.2.5.2 NARROWBAND DATA
For narrowband data the two alternatives described for the broadband
data are used. The first is to fit and subtract a smooth ring model
from each object exposure using MSCPUPIL. This is the same as described
for removing the pupil image from the flat field except the "type"
parameter is set to difference. The second is to create a sky flat
from disregistered exposures, extract the pupil pattern with MSCPUPIL,
and then subtract it from each object exposure using RMPUPIL.
4.2.6 DARK SKY OR TWILIGHT SKY FLAT FIELDS
You will notice that there are two flat field corrections which can be
performed by CCDPROC. The first one is for an initial flat field such
as the dome flat obtained at the beginning of the night, a standard
flat field from a previous night or run, or a final combined dome flat
and sky flat from some other night or run. The second is for a dark
sky or twilight sky flat field prepared from the object exposures after
they have been calibrated with the first flat field.
Sky flat fields are created by combining object exposures with objects
removed by using data in each pixel that is only sky. In principle one
could use exposures of the twilight sky but our experience is that
these do not work well. You are welcome to take some exposures and try
using them. We have found that dark sky flat fields derived from the
object exposures do work quite well.
Mosaic observations already typically dither a field. One will do even
better by combining observations from other fields. The more data used
the better the resulting sky flat will be. The main criterion for
including data is to avoid observations contaminated by varying
background light from the moon or scattered light from bright stars off
the field. Of course, another factor that has to be considered is
whether a field has a very large extended object which appears in many
of the observations. These will not be useful.
The sky flat field is created using the task SFLATCOMBINE with
parameters selected to reject objects appearing above a median. We
don't have much experience with creating sky flats currently so some
experimentation with parameters may be required. Below is one possibly
set of parameters.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = sflatcombine
input = List of images to combine
(output = Sflat) Output sky flat field root name
(combine= average) Type of combine operation
(reject = avsigclip) Type of rejection
(ccdtype= object) CCD image type to combine
(subsets= yes) Combine images by subset parameter?
(scale = mode) Image scaling
(statsec= ) Image section for computing statistics
(nkeep = 1) Minimum to keep (pos) or maximum to reject (neg)
(nlow = 1) minmax: Number of low pixels to reject
(nhigh = 1) minmax: Number of high pixels to reject
(mclip = yes) Use median in sigma clipping algorithms?
(lsigma = 6.) Lower sigma clipping factor
(hsigma = 3.) Upper sigma clipping factor
(rdnoise= rdnoise) ccdclip: CCD readout noise (electrons)
(gain = gain) ccdclip: CCD gain (electrons/DN)
(snoise = 0.) ccdclip: Sensitivity noise (fraction)
(pclip = -0.5) pclip: Percentile clipping parameter
(blank = 1.) Value if there are no pixels
(grow = 3.) Radius (pixels) for neighbor rejection
This task is a combination of CCDPROC to first process the images, if
they have not previously been processed, and COMBINE to combine the
offset images with rejection of object pixels. A new feature of this
task is the "grow" parameter which now provides a two dimensional
circular rejection of pixels around pixels rejected by the rejection
algorithm. Whatever rejection algorithm is used it is likely that the
best results will be when the clipping sigmas are non-symmetric as
shown above. Note that a very low rejection threshold or very large
grow radius will make the task quite slow.
After producing a good sky flat that has no evidence of objects it may
be applied directly to the data by using it as the second flat field
correction.
ms> ccdproc <files> sflatcor=yes sflat=Sflat*
Note that the object exposures used in creating the sky flat will
already have been processed except for the application of the sky flat
so CCDPROC will only apply the sky flat field calibration.
The sky flat field includes corrections at all scales from
pixel-to-pixel sensitivity variations to large scale illumination
differences. If the signal-to-noise is poorer than the dome flat field
you might wish to apply a filtering/smoothing operation to the sky flat
data thus relying on the dome flat field for the pixel-to-pixel
sensitivity calibration and the sky flat field for larger scale
illumination corrections. There are a number of filtering tasks in
IRAF. A median is a good filter and there is the choice of a ring
median or box median. To apply one of these general filtering tasks
you would use MSCCMD to run it on all the CCDs
ms> msccmd
msccmd: median $input $output 10 10
Input files: SflatV
Output files: SflatMedV
msccmd: q
Because the object exposures are first processed with the dome flat (or
other flat field) you would normally run CCDPROC again on the data
using the sky flat and any observations that have not been processed at
all will use both the dome flat and the sky flat. However, if you want
to make a single flat field to apply to raw data, say if starting over
or using it for a second night, you can combine the two flat field
corrections into a single flat field to be used as the only flat field
correction. This is done by multiplying the two flat fields using
MSCARITH
ms> mscarith FlatV * SflatV FinalflatV
4.2.7 THE VARIABLE PIXEL SCALE AND ZERO POINT UNIFORMITY
A key assumption in the traditional reduction of CCD images is that the
pixel scale is uniform and that a properly reduced blank sky image will
have a uniform and flat appearance. Unfortunately, this is not correct
when the pixel scale varies over the field. In the case of the NOAO
Mosaic at the Mayall telescope, the pixel scale decreases approximately
quadratically from the field center, with the pixels in the field
corners being 6% smaller in the radial direction, and 8% smaller in
area. Pixels in field corners thus would properly detect only 92% of
the sky level seen in the field center, even with uniform sensitivity.
At the same time the same number of TOTAL photons would be detected
from a star regardless of how many pixels the PSF would be distributed
over. Forcing the sky to be uniform over the image has the deleterious
effect of causing the photometric zeropoint to vary from center to
field corners by 8%. Note that this effect is different from
vignetting where the flux actually delivered to the image margins is
less than that at the center, an effect that IS corrected by the flat
field.
In practice, the photometric effect of the variable pixel scale can be
ignored provided that the reduced images will be part of a
dither-sequence to be stacked later on. As discussed below, prior to
stacking the images they first must be re-gridded, which produces
pixels of essentially constant angular scale. This is done with the
MSCIMAGE task, which re-grids the pixels and has a "flux conservation"
option that can scale the pixels photometrically by the associated area
change. If this function is disabled, then "improperly" flattened
images will have a uniform zero point restored. In short, the flat
field adjusted (if inappropriately) for the different pixel sizes, so
MSCIMAGE would then do no further adjustment. Stars would be too
bright in the corners of the flattened images, but after re-gridding,
their total fluxes would be seen to be scaled down to the appropriate
values.
If the mosaic CCD images are to be analyzed individually, as might be
done for standard star fields, then after the flat field reductions are
complete the differential scale effects must be restored. At present
we are developing a routine in the MSCRED package to do this, without
actually re-gridding the image. The correction process is simple; the
scale at any point in the Mosaic field is already known from the
astrometry so one just calculates and multiplies by the correction.
The final image would appear to have a variable sky level, but would be
photometrically uniform.
4.3 COORDINATE CALIBRATION
For some projects the basic flux calibrated CCD exposures may be all
that is required. However, if you want to obtain coordinate
information or combine multiple exposures which are dithered on the sky
or taken with different filters, you must calibrate the celestial world
coordinate system (WCS) of the data. This may be done in an absolute
or relative sense; an absolute calibration ties the data coordinates to
catalog coordinates while a relative calibration ties multiple
exposures to the same coordinates.
Determining the WCS from scratch is a complicated business and requires
special observations of astrometry fields. However, for NOAO Mosaic
data a standard coordinate calibration determined earlier is
automatically inserted into your data by the data capture agent. The
default coordinate system is sufficiently accurate for most purposes
and just requires some small adjustments as described below. To piece
a single exposure into a single image that does not require
registration to any other data you may use the default WCS and skip the
WCS calibration steps.
The WCS is a mapping from pixels in the mosaic data to celestial
coordinates relative to a reference point on the sky. The reference
point, or zero point, is set using the telescope pointing coordinate.
The telescope pointing is generally off by a small amount, though it
could be completely wrong in some hardware/software error situations.
In addition, differential atmospheric refraction introduces small axis
scale changes and rotations, which are significant due to the large
field of view of the mosaic even during the course of single set of
dithered exposured. Putting observations from different filters onto
the same coordinate system also requires mapping small scale changes,
since currently there is only a single standard WCS solution derived
through one filter. [In the future filter dependent solutions will be
made available.]
The WCS calibration operations consist of adjusting the standard
coordinate system calibration to a desired zero point and applying
small axis scale changes and rotations. This is done using objects
(usually stars) in the exposures. Unlike a full WCS calibration, which
requires a high density of stars with accurate catalog coordinates, the
adjustments to the default WCS calibration require only a few objects;
only one objects is needed to provide a zero point correction.
The WCS adjustments are determined by specifying coordinates for one or
more objects in the data. The coordinates can be obtained from a
reference catalog or, more commonly, by measuring coordinates from one
reference exposure to which other exposures are to be "registered". A
combination of using a catalog coordinate for one object in the field
to set the zero point in a reference exposure and then measuring the
positions of other stars in the reference image based on that zero
point calibration may also be done.
The two tasks you will use are MSCZERO and MSCCMATCH. MSCZERO is used
to interactively set the zero point of the coordinates, register
multiple exposures closely, and generate a list of coordinates in a
reference exposure to which other exposures in a dither set are
registered. MSCCMATCH finds objects at the positions specified by a
list of coordinates and determines corrections for the zero point, axis
scale change, and axis rotation.
4.3.1 SETTING COORDINATE ZERO POINTS AND MEASURING COORDINATES
MSCZERO is an interactive display task for mosaic exposures that allows
measuring coordinates and adjusting the WCS zero point. The task
parameters are shown below. The last set of parameters (starting with
"ra") are for the task to query and maintain lists.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = msczero
images = List of mosaic exposures
(nframes= 2) Number of frames to use
(logfile= default) Log file for measurements
ra = RA (hours)
dec = DEC (degrees)
update = yes Update WCS zero point?
(fd1 = )
(fd2 = )
The task displays each exposure in the list, in turn, and responds to
cursor key commands. You can go forward and backward through the input
list or quit at any point. The exposures are displayed by cycling
through the specified number of frames starting with the first frame.
As an aid to efficiency, if the exposure is already loaded in the
appropriate frame then the display step is skipped.
This task has several uses (type '?' to get the list of command
options):
1. Set the WCS zero point by specifying the coordinate of a star.
2. Create a list of coordinates for use with MSCCMATCH and MSCIMATCH.
3. Report coordinates at the cursor position.
It may be that the WCS zero points, based on the telescope pointing
coordinates, are accurate enough that you can use this task on only a
reference exposure to generate a list of coordinates for use with
MSCCMATCH and MSCIMATCH. However, because it is fairly quick to
explicitly check and set the zero point of all the exposures in a dither
set to the same coordinate for a common reference star, it is
recommended you do this first.
To check and set the zero points for a set of dithered exposures run
MSCZERO with a list of the exposures
ms> msczero @field1
After the first exposure is displayed either find a reasonably bright
unsaturated star which will be in all the exposures or find a star whose
coordinate is known from a catalog such as the HST Guide Star Catalog.
Move the cursor to the star and type 'z' (zero) to invoke a centering
algorithm. Note that even though the exposure may be displayed at
lower resolution the centering is done with the full resolution data.
The task will then tell you what it thinks the coordinate is and ask
you for a new coordinate. The first time 'z' is typed it will prompt
with the measured coordinate and thereafter it will prompt with the
last entered value. If you are referencing all the exposures to the
first exposure in the list accept the measured coordinate (and write
the value down in case you need it later) otherwise enter the desired
coordinate.
Note that all further measurements of the image will automatically
apply the measured zero point correction but the exposure WCS is not
actually updated until you type 'n' (next) or 'q' (quit). If you want
to print coordinates without changing the zero point correction use the
space bar or 'c' (center) to center on an object and print the centered
coordinate.
If you changed the WCS zero point you will be shown the zero point
offsets and given the option to update the WCS in the data file when
you type 'n'. Then the next exposure in the list will be displayed.
Find the same star and type 'z' again. Since it will retain the last
entered coordinate you should only need to accept the prompted
coordinates. When you have done this for all the exposures their
coordinate systems will be registered at least at that point.
The WCS in the dither set may still not be registered over all the
field due to refraction effects. Also the intensity scales of the
dithered exposures may not be the same due to changes in transparency
and sky brightness. These effects are calibrated by matching objects
throughout the field in position and brightness. This requires a list
of coordinates tied to one of the dithered exposures as a reference.
Usually the first exposure in the set is used as the reference. MSCZERO
is used to create a list from objects in the reference exposure.
ms> msczero obj021
Select objects, usually stars, throughout the field and type 'x' for
each one. This will center on the object and and record the coordinate
in a logfile. The default logfile name "default" creates a log file
beginning with "Coords." and followed by the name of the exposure. In
the example this will be "Coords.obj021".
To be useful for coordinate matching this list should have a good
number of stars, say three or four from each CCD, with emphasis on the
field edges but allowing for the dithering. For the intensity matching
you want to have stars with a range of brightness (though not saturated
or extremely faint) and which are mostly isolated so that a region
around them may be used for sky. The lists for the coordinate and
intensity matching do not have to be the same but it is reasonable to
just create one list.
4.3.2 MATCHING COORDINATE SYSTEMS
The task MSCCMATCH determines and applies a linear correction to the
WCS to match objects, generally stars, in an exposure to a set of
reference celestial coordinates. This correction maintains the
detector geometry and optical distortions while adjusting for changes in
apparent sky position such as produced by atmospheric refraction and
telescope pointing errors. The linear correction consists of a zero
point shift, scale changes in the right ascension and declination axes,
and rotations of the axes.
To use this task you need a list of reference celestial coordinates,
right ascension in hours and declination in degrees, and the mosaic
exposure coordinate system must be relatively close to the reference
coordinate system. The default WCS plus telescope pointing may be
close enough, but if not you would use MSCZERO to register the zero
points at some point in the exposures. Since it is relatively simple
to register a set of dithered exposures to a common star with MSCZERO
this is recommended procedure before using MSCCMATCH.
The reference coordinates should cover all of the mosaic field of view
to be sensitive to the small rotation and scale effects. The
coordinate list might be obtained from a catalog or measured from one
of the exposures to which other overlapping exposures will be matched.
For the purposes of making a well aligned stacked image from a set of
dithered exposures one generally uses one of the exposures as the
source of the reference coordinates.
MSCCMATCH operates on a set of input mosaic exposures; each in turn.
For an exposure it converts each input celestial coordinate to a pixel
coordinate in one of the extensions using the current WCS. If the
coordinate does not fall in any extension the coordinate is not used.
The pixel coordinate is used as a starting point for the APPHOT.CENTER
task. If the centering fails for some reason, such as the object being
too near the edge or the final position being too far from the initial
position, the coordinate is not used. For those objects successfully
found a fit is made between the original celestial coordinates and the
measured coordinates expressed as arc seconds from the exposure tangent
point. The fit is constrained to yield some combination of shift,
scale change, and rotation for each of the celestial coordinate axes.
These parameters are then used to update the exposure WCS so that the
adjusted measured coordinates best agrees with the reference
coordinates.
The task parameters are shown below.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = msccmatch
input = List of input mosaic exposures
coords = Coordinate file (ra/dec)
(nfit = 4) Min for fit (>0) or max not found (<=0)
(rms = 2.) Maximum fit RMS to accept (arcsec)
(maxshif= 5.) Maximum centering shift (arcsec)
(fitgeom= rxyscale) Fitting geometry
(update = yes) Update coordinate systems?
(interac= yes) Interactive?
(fit = yes) Interactive fitting?
(verbose= yes) Verbose?
accept = yes Accept solution?
The input is a list of mosaic exposures and a file of reference
celestial coordinates. The exposures should all include a significant
number of objects from the list of coordinates.
The task can be run interactively or non-interactively based on the
"interactive" parameter. In interactive mode you can graphically
interact with the fitting (selected with the "fit" parameter) and
accept or reject a fit based on the printed fit parameters. The
fitting is done using the task GEOMAP and the interactive mode allows
you to view the distribution of coordinates, residuals verses the input
coordinates, delete bad values, and possibly change the fitting
constraints (see the help for GEOMAP for more information).
The linear transformation may be constrained by the "fitgeometry"
parameter as described in the help for GEOMAP. This may be desirable
if there are only a few coordinates or if you want to impose some
physical assumption. Note that the effects of atmospheric refraction
actually do cause independent scale changes and rotations in the two
axes so the default "rxyscale" should be used.
There are some constraints which are placed on the task. The
"maxscale" parameter limits how far the objects may be found from the
initial coordinates. This constraint protects against incorrect
identifications and tells the centering routine how much of the image to
look at. This parameter should be as small as possible consistent with
the errors in the WCS. If you first zero the coordinates then the
objects should be found quite close to the initial coordinates. When
the "verbose" parameter is set the results of the centering will be
printed consisting of the image extension name, the final pixel
coordinates, the shift in pixel coordinates from the initial value, and
the formal uncertainties in the pixel coordinates. If an error occurs
one of the error codes from APPHOT.CENTER will be reported such as
"BigShift" for objects with too big a shift from the initial position
and "EdgeImage" for objects to near the edge of the image.
The "nfit" parameter requires a certain number of coordinates to be
included in the fit. If specified as a negative number the parameter
is interpreted as a maximum number that may be lost from the input list
due to being off the exposure or failing to be centered. The "rms"
parameter requires that the final RMS of the residuals about the fit
for each axis be less than a certain value.
4.4 PUTTING THE PIECES TOGETHER
This section tells you how to make single images from each
multiextension exposure and how to combine sets of dithered images into
a final deep image free from gaps and artifacts. Obtaining good
results depends on having well-flattened data, a uniform sky, a dither
pattern that samples the gaps and bad regions of the detectors, and
accurately registered world coordinates. Most difficulties are caused
by variable sky conditions or scattered light within a dither sequence
or the data used to create a sky flat.
4.4.1 REMOVING SKY GRADIENTS
Any sky level mismatches when combining dithered exposures produce
artifacts in the final image. The three sources of such mismatches are
sky gradients, sky level differences between the CCDs, and sky level
differences between exposures. While the flat field calibration,
particularly with a sky flat, should remove differences in sky levels
between CCDs, in practice there may still be small errors. And the flat
field will not deal with sky gradients across the large field of view.
Exposure-to-exposure sky brightness variations can be dealt with at a
later stage but even this is tricky.
The best final result is obtained by fitting a low order surface (a
plane or quadratic) to the sky and subtracting it from each CCD of each
object exposure at this stage. This will force the sky to be zero for
all CCDs and all exposures. Note that if one wants to preserve a sky
level for statistical reasons it is possible to add a uniform constant
after the subtraction to all the data (or add the constant to the final
dither stacked image).
To fit and subtract a sky and sky gradient the combination of IMSURFIT
and MSCCMD is used. With IMSURFIT use the option to fit to medians in
large blocks to remove the effects of objects.
ms> msccmd
msccmd: imsurfit $input $output xo=2 yo=2 type=resid xm=100 ym=100
Input files: obj*
Output files: obj*
msccmd: q
In this example the input and output are the same, replacing the
original by the sky subtracted data, but one can create new output
files if desired. Note that x and y orders of 2 correspond to a plane
and orders of 3 correspond to a quadratic surface.
4.4.2 CONSTRUCTING SINGLE IMAGES
Making a single image from a mosaic exposure is done by mapping the
pixels from each extension to a single uniform grid on the sky. The WCS
calibrations described in previous sections provide this. For making a
single image from a single exposure the WCS calibration is not critical
and the default WCS is sufficient. For combining multiple dithered
exposures all the exposures must be registered to a common coordinate
system, either relative to one reference exposure or to a set of catalog
stars, and each exposure must be resampled to the same final coordinate
system.
The task that makes single images from mosaic exposures is MSCIMAGE.
Its parameters are shown below.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mscimage
input = List of input mosaic exposures
output = List of output images
(referen= ) Reference image
(pixmask= yes) Create pixel mask?
(verbose= )_.verbose) Verbose output?
# Resampling parameters
(blank = 0.) Blank value
(interpo= linear) Interpolant for data
(minterp= linear) Interpolant for mask
(fluxcon= no) Preserve flux per unit area?
(ntrim = 7) Edge trim in each extension
(nxblock= 2048) X dimension of working block size in pixels
(nyblock= 1024) Y dimension of working block size in pixels
# Geometric mapping parameters
(interac= no) Fit mapping interactively?
(nx = 10) Number of x grid points
(ny = 20) Number of y grid points
(fitgeom= general) Fitting geometry
(functio=chebyshev) Surface type
(xxorder= 4) Order of x fit in x
(xyorder= 4) Order of x fit in y
(xxterms= half) X fit cross terms type
(yxorder= 4) Order of y fit in x
(yyorder= 4) Order of y fit in y
(yxterms= half) Y fit cross terms type
An output image is created for each input mosaic exposure. The output
image is created with a coordinate system defined by the specified
"reference" image. If no reference image is specified then the first
input mosaic exposure is used to define a simple tangent plane
coordinate system with optical distortions removed, and that coordinate
system is used for all the input mosaic exposures. The important point
is that for a set of dithered exposures all the output images must be
created with the same coordinate system grid so that they may be
combined by simple integer shifts along the image axes.
The normal usage is to specify all the mosaic exposures in a dither set
as the input, give a matching list of output images, and leave the
reference image unspecified. If all the exposures in a dither set are
not done at the same time then you must specify one of the earlier
output images as the reference image to continue to create the output
images on the same coordinate grid.
The output images are created with a size that just covers the input
data and initially filled with the specified "blank" value. This is the
value that the mosaic gaps will have in the final output image. Then
each extension is resampled into the appropriate part of the output
image. The coordinate mapping is generated by GEOMAP using the
geometric mapping parameters which you don't need to change. The
resampling is done with the specified interpolation function. The
small rotations in the CCDs produce edge effects in the interpolated
output pieces so a small trim is required to eliminate these. [At the
time this document was prepared the best value for the new science
grade NOAO Mosaic had not been determined.]
Linear interpolation is the fastest and most straightforward. Other
interpolation functions are available. In particular sinc
interpolation is now available as an add-on option (see the MSCRED
installation instructions). Experience with sinc interpolation shows
that it is not overly slow and does provide improved results;
particularly with maintaining the statistical characteristics of the
sky noise. The "minterpolant" parameter allows using a faster and more
local interpolation function for the mask. This is particularly useful
when using sinc interpolation of the data to allow flagging only around
the actual bad pixels and not extending out as far as the sinc
interpolation does.
It is useful for the later combining step to make bad pixel masks that
reflect the interpolation and resampling from the input data. These
may be created by setting the "pixmask" parameter. If this parameter
is set and the input mosaic data have bad pixel masks defined through
the header BPM keywords (default bad pixel masks are provided in the
NOAO Mosaic data) then the masks will be interpolated in exactly the
same way as the data. The interpolated masks will appear in the
working directory with names related to the output image names and with
the output images containing the BPM keyword pointing to these masks.
The input bad pixel masks are assumed to have zero for good data and
one for bad data and the output masks have zero for good data and
values between zero and ten thousand for bad data. The value is the
result of interpolation and reflects the relative contribution of good
and bad data.
The "fluxconserve" parameter applies a pixel area correction if
selected. As discussed earlier, standard flat fielding distorts the
flux per unit area in pixels of different projected size by making them
have the same flux per pixel. In effect this applies half of the flux
conservation operation by adjusting the pixel values without adjusting
the pixel sizes. MSCIMAGE does the second half by adjusting the pixel
sizes. So for standard flat fielded data, the usual route to making a
combined dithered image, the flux conservation parameter should not be
used to arrive at a proper final flux per unit area in the resampled
data. Flux conservation would only be used if the input mosaic data
has previously been corrected back to proper flux per unit area through
adjustment of the flat field or data for the variable pixel size
inherent in the mosaic coordinate system.
Below are two examples; one using prepared @files and one illustrating
advanced usage of filename templates.
ms> mscimage @dither1 @outdither1 pixmask+
ms> mscimage obj02![2-5]* %obj%mos%02![2-5]* pixmask+
In the second example the input template expands to obj022.fits to
obj025.fits and the output template matches the input template using the
first part of the %% substitution field and then replaces the "obj" with
"mos" to give output images mos022.fits to mos025.fits.
4.4.3 MATCHING INTENSITY SCALES
When stacking dithered exposures (the single images created in the
previous step) to fill in the mosaic gaps and remove bad pixels and
cosmic ray events it is critical that the intensity scales of the
images match. Otherwise you will see artifacts from the gaps, places
with bad data, and around objects as the combined intensity level jumps
when data from an exposure is missing or rejected. Also the rejection
algorithms require that the image intensities match both at the sky
level and in the objects.
There are two parameters that must be determined to match the intensity
scales. One is a additive offset caused by sky brightness variations.
The second is a multiplicative scale change caused by transparency and
exposure time variations. Matching the intensity scales for a set of
dithered exposures consists of determining values for these two scaling
parameters relative to a reference exposure and setting them in the
image headers. The actual adjustment of the pixels values occurs when
stacking the exposures.
The intensity matching values are determined by the task MSCIMATCH.
The task parameters are shown below.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mscimatch
input = List of images
coords = File of coordinates
(scale = yes) Determine scale?
(zero = no) Determine zero offset?
(box1 = 21) Inner box size for statistics
(box2 = 51) Outer box size for statistics
(lower = 1.) Lower limit for good data
(upper = INDEF) Upper limit for good data
(niterat= 3) Number of sigma clipping iterations
(sigma = 3.) Sigma clipping factor
(interac= no) Interactive?
(verbose= yes) Verbose?
The input is a list of images to be matched and a file of celestial
coordinates (RA in hours and DEC in degrees) to use in computing the
matching parameters. The input images are the single images constructed
from the mosaic exposures for a set of dithered observations.
The parameters "scale" and "zero" select whether to determine the
multiplicative scale, the zero level offsets, or both. If the sky has
been subtracted at an earlier stage (as recommended) then only the
multiplicative scale difference needs to be determined. The advantage
of subtracting the sky earlier is that scale determination becomes
better constrained. Also determining the sky from photometry (as done
by this task) is less robust than the surface fitting which uses all of
the sky data.
The scaling parameters are determined by measuring the mean flux in a
set of matching regions between each input image. The centers of the
regions are specified by their celestial coordinates. The list of
coordinates should consist of the positions of objects in the field.
These objects should span a range of brightness. As noted previously,
you would normally use the same coordinate list as used with MSCCMATCH,
which is generally obtained using MSCZERO. However, you can use any
IRAF task that produces a list of celestial coordinates from images
with a WCS. One possibility is to use RIMCURSOR on one of the
displayed single images with the "wcs" parameter set to "world" and the
"wxformat" set to "%.2H" to produce right ascension values in hours
instead of degrees.
The now accurately aligned coordinate systems in the images are used to
identify the matching pixel coordinate center in each image. The
regions to be measured consist of square boxes of the specified sizes
about the pixel coordinate center. There are two boxes, an inner box
and an outer box which excludes the inner box. The box sizes are
intended to define photometry apertures for the objects and nearby
background. It is not critical that they exactly fit the objects or
that the objects necessarily be stars but this is usually how they will
be set. Because of possible PSF variations the inner box should be
large enough include all the light from stars over the whole data set.
If the inner box is not fully contained in the input or reference image
that box is not used for that pair. Similarly the outer box must be
fully contained in the images but if only the outer box is outside one
or both images the measurement for the inner box may still be used.
In order to exclude regions that include the gaps or bad data in one or
both of the pair of images all pixels in a box must have values between
the specified good data limits. Those regions with values outside the
limits are eliminated from the intensity matching.
The mean fluxes in each region are used to simultaneously fit the
relations
mean_j = A_ij + B_ij * mean_i
for all i and j where i and j are any pair of images. These equations
are constrained by the fact that the scaling from image i to j,
followed by the scaling from image j to k, must agree with the scaling
from image i to image k. The final scaling coefficients reported and
stored in the image header are A_1j and B_1j, which correspond to the
scalings to the first image in the input list.
The task will attempt to reject photometry points which are discrepant.
If the task is run interactively it will also show plots of the
photometry flux in one image verses another. It does this for
sequential pairs of images. Points can be deleted in these plots and
they will be excluded from the data used to determine the scaling
parameters.
When the task is done determining the scaling factors they will be
printed and a prompt issued to accept or not accept the results. If
the scaling parameters are accepted then the keywords MSCZERO and
MSCSCALE are recorded in the input image header when the "update"
parameter is set. Note that the reference image is assigned values of
0 and 1 for these header keywords.
4.4.4 MAKING THE FINAL STACK IMAGE
After MSCIMAGE produces single images of each of the dithered mosaic
exposures with a common coordinate system grid, a final image is created
with the task MSCSTACK. The task MSCIMATCH is generally used to match
the intensity scales of the images before this step as described in the
previous section. However, for quick reductions or for other reasons
the images may be stacked either with no intensity matching or using
the "scale" and "zero" options of MSCSTACK. The task parameters are
shown below.
I R A F
Image Reduction and Analysis Facility
PACKAGE = mscred
TASK = mscstack
input = List of images to combine
output = Output image
(plfile = ) List of output pixel list files (optional)
(combine= median) Type of combine operation (median|average)
(reject = none) Type of rejection
(masktyp= none) Mask type
(maskval= 0.) Mask value
(blank = 0.) Value if there are no pixels
(scale = !mscscale) Image scaling
(zero = !msczero) Image zero point offset
(weight = none) Image weights
(statsec= ) Image section for computing statistics
(lthresh= 1.) Lower threshold
(hthresh= INDEF) Upper threshold
(nlow = 1) minmax: Number of low pixels to reject
(nhigh = 1) minmax: Number of high pixels to reject
(nkeep = 1) Minimum to keep (pos) or maximum to reject (neg)
(mclip = yes) Use median in sigma clipping algorithms?
(lsigma = 3.) Lower sigma clipping factor
(hsigma = 3.) Upper sigma clipping factor
(rdnoise= 0.) ccdclip: CCD readout noise (electrons)
(gain = 1.) ccdclip: CCD gain (electrons/DN)
(snoise = 0.) ccdclip: Sensitivity noise (fraction)
(sigscal= 0.1) Tolerance for sigma clipping scaling corrections
(pclip = -0.5) pclip: Percentile clipping parameter
(grow = 0) Radius (pixels) for 1D neighbor rejection
This task is a simple variant of COMBINE that registers the images
using the coordinate systems and has the default threshold parameters
set to ignore values below one DN based on the default "blank" value in
MSCIMAGE for the gaps. If you have also generated bad pixel masks for
the resampled images you can exclude them as well by setting "masktype"
to "goodvalue".
The real art in using this task is deciding how to scale and reject bad
data not covered by the bad pixel masks. A "combine" of "median" is
the simplest but it does not optimize the signal-to-noise for the
number of images. If you "average" the data you will probably want to
apply a rejection algorithm such as "avsigclip".
Careful flat fielding will make each separate image have the same sky
level across the different CCDs. However, the sky levels and
transparency may still be varying from exposure to exposure. If you
simply combine such data you will see imprints of the gaps. So it is
generally a good idea to scale the images. This is done using the
"scale" and "zero" parameters which can be set to header keyword values,
files containing the values, or special values to compute image
statistics in a particular region of the data. The recommended method
for scaling is to use the intensity matching task MSCIMATCH described
in the previous section and use the image header keywords MSCSCALE and
MSCZERO produced by that task.
An example of using this task to create a final image is given below.
ms> mscstack @field1 Field1 combine=average rej=avsigclip
IRAF Project
National Optical Astronomy Observatories