MODULE: argus, ctioslit, hydra, iids, irs, kpnocoude, kpnoslit
SYSTEM: V2.12a-V2.12b
NUMBER:	547 (see also 544)
MODULE:	argus, ctioslit, hydra, iids, irs, kpnocoude, kpnoslit
SYSTEM:	V2.12a-V2.12b
DATE:	Mon Apr 25 10:59:24 MST 2005
FROM:	valdes

BUG:    Changes to SCOMBINE in V2.12.2a made it a separate executable.  The
	package definitions for the packages listed above (see also number
	544 concerning the ECHELLE package) failed to be updated so when
	trying to run SCOMBINE (or a task that calls SCOMBINE such as
	DOHYDRA) the following error message is produced.

            ERROR: task `scombine' has no param file

	The solution is editing the $.cl file, where
	 is one of the modules listed above,  so that it reads as
	shown below.  A simpler workaround is to load the onedspec package
	and run scombine from that package.

STATUS: Fixed for the next release.

Modifications to $.cl:

...
task    ...,
	[scombine,]						<-- Delete line
	...,
	...		= onedspec$x_onedspec.e
task    scombine        = 'onedspec$scombine/x_scombine.e'	<-- Add line
...

Note that scombine has to be removed from the first block of task
definitions and added as a separate task statement as shown.


Comments (0)